EMC Requirements for Medical Devices

Steve Hayes EMC Expert
By Engaged Expert Steve Hayes

More contentfrom this author

Learn about ourEngaged Experts

EMC Requirements for Medical Devices

Medical devicesare some of the most heavily regulated product types available on the market today, and rightly so – any type of product failure can have negative consequences, from a patient not seeing the therapeutic benefit of the device to an injury or even a fatality. When the medical device is merged with a wireless functionality, the regulatory framework becomes even more complex and is subject to greater knowledge of regulatory compliance to ensure that the overall product is safe, fit for purpose and compliant with the various regulations. As a result,EMC testing of medical devicesis a critical part of any testing program.

Understandably, many medical device manufacturers focus on the clinical aspects of their product and don’t always understand the impact of its use in close proximity to other electrical equipment or wireless devices. This paper seeks to highlight these issues and provide guidance on what to do.

EMC Testing of Non-Radio Medical Devices

First, it is important to understand that in the medical world, the term ‘EMC’ is not sufficient to categorise the assessment of Electromagnetic Disturbances as required by the various regulations throughout the world. For EMC testing of medical devices, it is the basic safety that the product provides as well as the essential performance that need to be assessed to ensure that the expected benefit and protection that it is designed to provide. The requirements can be considered as being similar to functional safety for the assessment of the EM disturbances in other areas of industry.

EM testing is not, and cannot, be the sole responsibility of the EMC test engineer or community to define the EM requirements for a medical device since the clinical evaluation and risk management are essential to achieving both electromagnetic compatibility and electromagnetic safety. The actual EM testing is a relatively small part of medical device approval and takes into account the environment as well as EM risk identification, analysis and controls to mitigate the threats posed by other electrical equipment.

EMC Testing of Medical Devices

"Any type of product failure can have negative consequences, from a patient not seeing the therapeutic benefit of the device to an injury or even a fatality." Steve Hayes
IEC 60601 Test Methods

As any medical device manufacturer will be aware, the IEC updated the medical EMC testing standard,IEC 60601-1-2 in 2014, and it now stands at Edition 4. There is a fundamental difference between the current and early versions of the standard - it is now a medical device safety testing standard, where before it was a medical device EMC testing standard.

While many of the same methods are included in both versions, which at first glance makes the changes seem minor, emphasis has shifted from a standard based on functionality to one based on safety. To underline the differences, the IEC has issued a new document, IEC 60601-4-2, which provides greater clarity on the issue and the difference between electromagnetic performance and electromagnetic safety assessments.

Medical Device EMC Testing and Risk Management

Risk Management is at the very heart of EMC testing of medical devices, and a manufacturer is required to perform an assessment based on the requirements of the standard in conjunction with any risks they identify. Additionally, the Food and Drug Administration (FDA) in the US requires manufacturers to consider the implications when exposed to known RF devices used in medical centres such as hospitals. AIM 7351731 specifically addresses the threats from Radio Frequency Identification Readers (RFID) and is recognised by FDA for doing so. It applies to any medical electrical product that could be exposed to RFID and assesses the specific modulations (across a number of frequency bands on which they operate).

Using the experience that Element has for conventionalEMC testing, we have developed a range of services to help manufacturers prepare and test against this standard, ensuring that these types of risk are mitigated for EMC testing of medical devices.

无线测试的医生al Devices

When a medical electrical product incorporates a wireless function, it increases the regulatory requirements considerably. In addition to supplementary EMC assessments, manufacturers need to consider the international transmissions from the product to ensure that the product operates on the correct frequency, transmits the right power level, and other essential radio characteristics. These aspects are regulated through agencies such as theFCC(US),ISED(Canada), and the EuropeanRadio Equipment Directive(CE marking).

Regarding the EMC for awireless device, the manufacturer needs to determine if the presence (or lack of) a wireless link affects the essential performance or basic safety of the device. Clearly if a wireless device cannot communicate, its performance will be affected and its intended functionality is undermined.

鉴于越来越多的产品设计wireless use, the radio spectrum is becoming very congested, increasing the probability of interference– Bluetooth, Wi-Fi, ZigBee , and ANT+ are all technologies that use the 2.4GHz ISM band and must coexist with one another to ensure they work as intended.

The loss of Wi-Fi for a smartphone would be a mere inconvenience, until the phone itself is used as part of a medical decision process to speed up an essential evaluation process. Remote alarms that rely on a Wi-Fi connection to a nurse station could equally have a major impact if it were being used to alert critical signs of a patient or IV drug failure. Appropriate radio interference and immunity testing of medical devices helps to ensure that these critical functions continue to perform in an increasingly crowded space.

“Given that more and more products are designed for wireless use, the radio spectrum is becoming very congested, increasing the probability of interference.”

Medical Device Radio Testing and Risk Management

The Association for the Advancement of Medical Instrumentation (AAMI) published a standard (AAMI TIR69:2017 - Risk management of radio-frequency wireless coexistence for medical devices and systems) to help guide manufacturers with the assessment of wireless devices, ensuring that they coexist with other wireless products. The document is used in conjunction with ANSI C63.27:2017 (American National Standard for Evaluation of Wireless Coexistence), which provides the test methods and is an essential part of ensuring that a wireless device is fit for purpose in its intended EM environment and will work alongside other wireless devices.

The assessment for coexistence involves the communication of the medical product, followed by the introduction of interfering signals close or harmonically related to the fundamental frequency. This ensures that the essential performance and communication of the product is maintained.

Element has invested heavily in a range of services to help, guide and test for wireless coexistence and is able to offer these services in a variety of locations both in the UK and US.